System capacity: 120 groups program X100 sections, required segmentation sections of each group can be arbitrarily divided and each program can freely link another.

Cycle setting: each operation procedure can carry out 9999 times or heaps of times cycle and repeated performance, and to the degree, divide further 5 segmentation sections to do part cycle of the extra.

Offer CE and ISO 17025 calibration certicate
Temperature And Humidity Controlled Cabinets Can custom-made according different request

Application

For conditioning of samples prior to testing. It also can be used for a variety of materials of high - low temperature alternating test. The test temperature, humidity and time can be programmed.

Control System

1. Touch screen, South Korean import,900 temperature controller

2. Control mode: Touchpad

3. Display: LCD display, Korean TEMi880, programmable.

4. System capacity: 120 groups program X100 sections, required segmentation sections of each group can be arbitrarily divided and each program can freely link another.

5. Cycle setting: each operation procedure can carry out 9999 times or heaps of times cycle and repeated performance, and to the degree, divide further 5 segmentation sections to do part cycle of the extra.

Key Specification

Model

ULB-E01-80

ULB-E01-150

Capacity

80L

150L

Inside dimension (WxHxD,cm)

40x50x40

50x60x50

Outside dimension (WxHxD,cm)

95x136x94

100X146X104

Temperature range

Regular: -20~150℃

Customized : -40~150℃ or -60~150℃

Humidity range

20%~98%, can be set.

 temperature humidity environmental chamber

Rate of temperature increasing

Room temperature ~150℃

About 45 minutes(non-linear no-load, about 3.5℃/minute

Room temperatur~ -20℃

About 45 minutes(non-linear no-load, about 1.2℃/minute)

Machine precision

Control accuracy (test precision)


Temperature

+/-0.5℃

Humidity

+/-1%RH

Distribution accuracy (uniformity)

Temperature

+/-1.0℃

Humidity

+/-2%RH

Environment temperature

5℃-28℃

Environment relative humidity

≤85%RH

Power voltage

220V AC(+/-10%), protect ground wire, ground resistance≤4Ω.

Frequency

50+/-0.5Hz

Power switch

Air-break switch(40A, knife switch or outlet is prohibited to use.)

Power

4.5KW

Humidification water

Resistivity(≥500MΩ)

Requirement of ambient

air quality

Do not contain high concentration dust and flammable gas, without strong electromagnetic radiation source nearby.  

Power supply

AC1¢3W 380V 50/60HZ, voltage change allowed value: ±10% of fixed voltage. 

Max loading capacity

8KVA

Max operating current

30A