China Temperature and Humidity Controlled Cabinets | ULB-E01 | Certified Factory Suppliers
Application
For conditioning of samples prior to testing. It also can be used for a variety of materials of high - low temperature alternating test. The test temperature, humidity and time can be programmed.
Control System
-
1
Touch screen: South Korean import, 900 temperature controller.
-
2
Control mode: Touchpad.
-
3
Display: LCD display, Korean TEMi880, programmable.
-
4
System capacity: 120 groups program X 100 sections, required segmentation sections of each group can be arbitrarily divided and each program can freely link another.
-
5
Cycle setting: Each operation procedure can carry out 9999 times or heaps of times cycle and repeated performance, and to the degree, divide further 5 segmentation sections to do part cycle of the extra.
Key Specification
| Model | ULB-E01-80 | ULB-E01-150 |
| Capacity | 80L | 150L |
| Inside dimension (WxHxD, cm) | 40x50x40 | 50x60x50 |
| Outside dimension (WxHxD, cm) | 95x136x94 | 100X146X104 |
| Temperature range |
Regular: -20~150℃ Customized: -40~150℃ or -60~150℃ |
|
| Humidity range | 20%~98%, can be set. | |
| Rate of temperature increasing | Room temperature ~ 150℃ | About 45 minutes (non-linear no-load, about 3.5℃/minute) |
| Room temperature ~ -20℃ | About 45 minutes (non-linear no-load, about 1.2℃/minute) | |
| Machine precision (Control accuracy) | Temperature | +/-0.5℃ |
| Humidity | +/-1%RH | |
| Machine precision (Distribution accuracy / uniformity) | Temperature | +/-1.0℃ |
| Humidity | +/-2%RH | |
| Environment temperature | 5℃-28℃ | |
| Environment relative humidity | ≤85%RH | |
| Power voltage | 220V AC(+/-10%), protect ground wire, ground resistance≤4Ω. | |
| Frequency | 50+/-0.5Hz | |
| Power switch | Air-break switch (40A, knife switch or outlet is prohibited to use.) | |
| Power | 4.5KW | |
| Humidification water | Resistivity (≥500MΩ) | |
| Requirement of ambient air quality | Do not contain high concentration dust and flammable gas, without strong electromagnetic radiation source nearby. | |
| Power supply | AC1¢3W 380V 50/60HZ, voltage change allowed value: ±10% of fixed voltage. | |
| Max loading capacity | 8KVA | |
| Max operating current | 30A | |
Frequently Asked Questions (FAQ)
What is the primary application of this environmental chamber?
This chamber is designed for conditioning samples prior to testing and conducting high-low temperature alternating tests on a variety of materials. It features programmable controls for temperature, humidity, and testing time.
What temperature and humidity ranges does the chamber support?
The regular temperature range is -20℃ to 150℃, with customized options available for -40℃ to 150℃ or -60℃ to 150℃. The configurable humidity range is 20% to 98% RH.
How advanced is the control system?
The chamber is equipped with a South Korean imported TEMi880 programmable LCD touch screen controller. It supports up to 120 program groups with 100 sections each, allowing complex segmentations and cycle repetitions up to 9999 times.
What are the power supply and installation requirements?
It requires an AC 1-phase 3-wire 380V (50/60Hz) power supply (voltage changes allowed up to ±10%) and a 40A air-break switch. Traditional knife switches or outlets are strictly prohibited. Protective ground wiring is required with a resistance of ≤4Ω.
What type of water is required for the humidification system?
The system requires high-quality humidification water with a resistivity of ≥500MΩ to ensure proper function and prevent mineral buildup.
What ambient conditions are required for the installation site?
The site must have an ambient temperature between 5℃ and 28℃ and relative humidity of ≤85% RH. The surrounding air must be free of high-concentration dust, flammable gases, and strong electromagnetic radiation sources.